10/12 IM MT300-71EN
:CALibrate:DA:POINt
Function Sets or queries the adjustment point of the D/A
output range to be adjusted.
Syntax
:CALibrate:DA:POINt {PFULL|MFULL
|ZERO}
CALibrate:DA:POINt?
PFULL: Full-scale positive value
MFULL: Full-scale negative value
ZERO: Zero point
Example
:CALIBRATE:DA:POINT PFULL
:CALIBRATE:DA:POINT?
-> :CALIBRATE:DA:POINT PFULL
Description • Onmodelswithoutthe/DAoption,a“241:
Hardwaremissing”errorwilloccur.
• Use the
:CALibrate:FUNCtion
command in
advance to set the adjustment target to DA. If
theadjustmenttargetisnotDA,a“221:Setting
conflict”errorwilloccur.Aqueryreturnsthe
current value.
:CALibrate:DA:RANGe
Function Sets or queries the D/A output range to be
adjusted.
Syntax
:CALibrate:DA:RANGe <
Voltage
>
:CALibrate:DA:RANGe?
<Voltage> = 2, 5 (V)
Example
:CALIBRATE:DA:RANGE 5V
:CALIBRATE:DA:RANGE?
-> :CALIBRATE:DA:RANGE 5.0E+00
Description • Onmodelswithoutthe/DAoption,a“241:
Hardwaremissing”errorwilloccur.
• Use the
:CALibrate:FUNCtion
command in
advance to set the adjustment target to DA. If
theadjustmenttargetisnotDA,a“221:Setting
conflict”errorwilloccur.Aqueryreturnsthe
current value.
:CALibrate:DA:WRITe
Function Writes the adjustment value of D/A output.
Syntax
:CALibrate:DA:WRITe
Example
:CALIBRATE:DA:WRITE
Description • Onmodelswithoutthe/DAoption,a“241:
Hardwaremissing”errorwilloccur.
• Use the
:CALibrate:FUNCtion
command in
advance to set the adjustment target to DA. If
theadjustmenttargetisnotDA,a“221:Setting
conflict”errorwilloccur.
• Whenexecuted,theadjustmenttargetissetto
NONE.
:CALibrate:DMM:ADJust:MFULl
Function Sets the full-scale negative adjustment value of
the voltage/current measurement.
Syntax
:CALibrate:DMM:ADJust:MFULl <NRf>
<NRf> =
5 V range
−4.00000(V)to−6.00000(V)(0.00001Vsetting
resolution)
20 mA range
−16.0000(mA)to−24.0000(mA)(0.00001mA
setting resolution)
Example
:CALIBRATE:DMM:ADJUST:MFULL−5.00000
Description • Onmodelswithoutthe/DMoption,a“241:
Hardwaremissing”errorwilloccur.
• Use the
:CALibrate:FUNCtion
command in
advance to set the adjustment target to DMM.
IftheadjustmenttargetisnotDMM,a“221:
Settingconflict”errorwilloccur.
• Use the
:CALibrate:DMM:POINt
command
to set the adjustment point to MFULL. If the
adjustmentpointisnotMFULL,a“221:Setting
conflict”errorwilloccur.
:CALibrate:DMM:ADJust:PFULl
Function Sets the full-scale positive adjustment value of
the voltage/current measurement.
Syntax
:CALibrate:DMM:ADJust:PFULl <NRf>
<NRf> =
5 V range
4.00000(V)to6.00000(V)(0.00001Vsetting
resolution)
20 mA range
16.0000(mA)to24.0000(mA)(0.00001mA
setting resolution)
Example
:CALibrate:DMM:ADJust:PFULl5.00000
Description • Onmodelswithoutthe/DMoption,a“241:
Hardwaremissing”errorwilloccur.
• Use the
:CALibrate:FUNCtion
command in
advance to set the adjustment target to DMM.
IftheadjustmenttargetisnotDMM,a“221:
Settingconflict”errorwilloccur.
• Use the
:CALibrate:DMM:POINt
command
to set the adjustment point to PFULL. If the
adjustmentpointisnotPFULL,a“221:Setting
conflict”errorwilloccur.