:CALibrate:DMM:DATE?
Function Queries the date when the adjustment value of
voltage/current measurement was written.
Syntax
:CALibrate:DMM:DATE?
Example
:CALIBRATE:DMM:DATE?
->:CALIBRATE:DMM:DATE"2019/09/12"
Description Onmodelswithoutthe/DMoption,a“241:
Hardwaremissing”errorwilloccur.
:CALibrate:DMM:INITialize
Function Initializes the adjustment value of voltage/current
measurement.
Syntax
:CALibrate:DMM:INITialize
Example
:CALIBRATE:DMM:INITIALIZE
Description Initializes the adjustment value of voltage/
current measurement. On models without the /
DMoption,a“241:Hardwaremissing”errorwill
occur.
Use the
:CALibrate:FUNCtion
command in
advance to set the adjustment target to DMM.
IftheadjustmenttargetisnotDMM,a“221:
Settingconflict”errorwilloccur.
:CALibrate:DMM:MEASure?
Function Queries the measured voltage/current.
Syntax
:CALibrate:DMM:MEASure?
Example
:CALIBRATE:DMM:MEASURE?
->:CALIBRATE:DMM:MEASURE5.0000E+00
Description • Onmodelswithoutthe/DMoption,a“241:
Hardwaremissing”errorwilloccur.
If the value is +OverRange or +OverFlow,
9.90E+37isreturned.
Ifthevalueis−OverRangeor−OverFlow,
−9.90E+37isreturned.
If DMM is not selected with the
:CALibrate:FUNCtion
command,9.91E+37
is returned.
:CALibrate:DMM:POINt
Function Sets or queries the adjustment point of the
measurement range of the voltage/current
measurement to be adjusted.
Syntax
:CALibrate:DMM:POINt {PFULL|MFULL}
:CALibrate:DMM:POINt?
PFULL: Full-scale positive value
MFULL: Full-scale negative value
Example
:CALIBRATE:DMM:POINT PFULL
:CALIBRATE:DMM:POINT?
-> :CALIBRATE:DMM:POINT PFULL
Description • Onmodelswithoutthe/DMoption,a“241:
Hardwaremissing”errorwilloccur.
Use the
:CALibrate:FUNCtion
command in
advance to set the adjustment target to DMM.
IftheadjustmenttargetisnotDMM,a“221:
Settingconflict”errorwilloccur.Aqueryreturns
the current value.
:CALibrate:DMM:RANGe
Function Sets or queries the measurement range of the
voltage/current measurement to be adjusted.
Syntax
:CALibrate:DMM:RANGe{VOLTage,
<
Voltage
>|CURRent,<
Current
>}
:CALibrate:DMM:RANGe?
<Voltage> = 5 (V)
<Current> = 20 (mA)
Example
:CALIBRATE:DMM:RANGEVOLTAGE,5V
:CALIBRATE:DMM:RANGE?
5 V range
->:CALIBRATE:DMM:RANGEVOLTAGE,
5.000E+00
20 mA range
->:CALIBRATE:DMM:RANGECURRENT,
20.0E−03
Description • Onmodelswithoutthe/DMoption,a“241:
Hardwaremissing”errorwilloccur.
Use the
:CALibrate:FUNCtion
command in
advance to set the adjustment target to DMM.
IftheadjustmenttargetisnotDMM,a“221:
Settingconflict”errorwilloccur.Aqueryreturns
the current value.
:CALibrate:DMM:WRITe
Function Writes the adjustment value of voltage/current
measurement.
Syntax
:CALibrate:DMM:WRITe
Example
:CALIBRATE:DMM:WRITE
Description • Onmodelswithoutthe/DMoption,a“241:
Hardwaremissing”errorwilloccur.
Use the
:CALibrate:FUNCtion
command in
advance to set the adjustment target to DMM.
IftheadjustmenttargetisnotDMM,a“221:
Settingconflict”errorwilloccur.
• Whenexecuted,theadjustmenttargetissetto
NONE.