6-21
IM 04P01B01-01E
Setup Operations for Convenient Functions (Setting Mode)
1
2
3
4
5
6
7
8
9
10
11
12
App
Index
MSG format
Choose a message format. Measured values,* message strings, and date/time can be
set arbitrarily.
13:H:M L16
CHR:Date/Time
CHR:01...
CHR:0A...
CHR:L01...
CHR:Date/Time
:
Measured value
Computed value (/M1 option)
Message string (number of characters)
Date/Time
Comment
CHARACTER key
Example: Based on the message 1 string of the standard function, you can set a
combination with message format 1 and print it out as message 1 including
measured values.*
10th and 11th character of the
message string (9 characters
already used, so this represents
the next characters)
Standard function
Message 1 string: Process-1°C
Message 1 printout result
: 06/30 10:10 Process-1 134.8°C
Message format 1
: M/D H:M L09 01L02
Measured value on CH1 (no units)
Space (number of characters: 1)
Space
1-9th characters of the message string
Date/time
* Computed values can also be included on models with the /M1 option.
Date/Time Format and Number of Characters Used
H:M (Hour:Minute) 5 characters
H:M:S (Hour:Minute:Second) 8 characters
M/D H:M (Month/Day Hour:Minute) 11 characters
M/D H:M:S (Month/Day Hour:Minute:Second) 14 characters
D/M H:M (Day/Month Hour:Minute) 11 characters
D/M H:M:S (Day/Month Hour:Minute:Second) 14 characters
D.M H:M (Day.Month Hour:Minute) 11 characters
D.M H:M:S (Day.Month Hour:Minute:Second) 14 characters
M.D H:M (Month.Day Hour:Minute) 12 characters
M.D H:M:S (Month.Day Hour:Minute:Second) 15 characters
Y/M/D H:M:S (Year/Month/Day Hour:Minute:Second) 19 characters
M/D/Y H:M:S (Month/Day/Year Hour:Minute:Second) 19 characters
D/M/Y H:M:S (Day/Month/Year Hour:Minute:Second) 19 characters
D.M.Y H:M:S (Day.Month.Year Hour:Minute:Second) 19 characters
M.D.Y H:M:S (Month.Day.Year Hour:Minute:Second) 20 characters
Format of Measured Values and Computed Values, and Number of
Characters Used
The format for measured values is 01, 02, ...06 (depending on the model). No units
are added. 7 characters are used.
The format for computed (/M1 option) values is: 0A, 0B, 0C, 0D, 0E, 0F, 0G, 0J, 0K,
0M, 0N, and 0P (depending on the model). No units are added. 9 characters are
used.
Character String Format
The format for message strings is L01 (1 character), L02 (2 characters), .... L16 (16
characters).
Intheexampleabove,L09species“Process-1”andL02species“°C.”
<Related Topics> Enabling the message format: Section 7.24
6.15 Setting the Message Format (/BT1 Option)