6-21
IM MV1000-01E
Display Configuration
1
2
3
4
5
6
7
8
9
10
11
12
13
App
Index
Settings
View > Partial
If you select On, the Partial item appears in the Setting Mode menu.
First-CH, Last-CH
Select the target channels.
Partial > On/Off
To enable the partially expanded display function, select On.
Partial > Expand
Set the position where the value specied by Boundary will be displayed as a
percentage of the display span from within the range of 1 to 99.
Partial > Boundary
Set the boundary value between the reduced section and the expanded section from
within the range of “the minimum span value + the smallest number expressible given
the decimal place” to “the maximum span value – the smallest number expressible
given the decimal place.” For channels that are set to scaling, the selectable range is
“the minimum scale value + the smallest number expressible given the decimal place”
to “the maximum scale value – the smallest number expressible given the decimal
place.”
Example:
Input range: –6 V to 6V. Expand: 30. Boundary: 0
The –6 V to 0 V range is displayed in the 0% to 30% range, and the 0 V to
6 V range is displayed in the 30% to 100% range.
6.9 Partially Expanding a Waveform