IM 12B6B3-E-H
Installation and wiring 3-9
Figure 3-9b. Impedance measurement jumper setting
4
3
Jumper connections for :
- Input 1, Low impedance
- Input 2, High impedance
Service code 03 = 0.0.1
- Input 1, Low impedance, Temp. comp. off, Check on
Service code 04 = 1.1.1
- Input 2, High impedance, Temp. comp. on, Check on
Jumper connections for :
- Input 1, Low impedance
- Input 2, Low impedance
Service code 03 = 0.0.1
- Input 1, Low impedance, Temp. comp. off, Check on
Service code 04 = 0.0.1
- Input 2, Low impedance, Temp. comp. off, Check on