EasyManua.ls Logo

YOKOGAWA Rotamass series - Connecting the connecting cable to transmitter

YOKOGAWA Rotamass series
90 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
General Instruction Manual
Wiring
Connecting cable installation
50 / 90
IM 01U10B00-00EN-R, 3rd edition, 2018-07-09
7.3.3 Connecting the connecting cable to transmitter
1. Loosen the four screws from the sensor connection cover and remove cover.
2. Remove connector nut from cable gland and pull out clamped insert.
3. Push cable through connector nut and clamped insert.
4. Remove precut outer casing of cable.
5. Pull back outer shield of cable over clamped insert.
6. Feed cable through cable entry into transmitter.
7. Use the operating tool to connect wires to connection terminals in accordance with
terminal diagram (Connection terminals of connecting cable).
8. Assemble the cable gland and tighten connector nut.
9. Place sensor connection cover onto transmitter and fasten with four screws.

Table of Contents

Related product manuals