IM 12D08N05-01E
2-3
<2. GENERAL SPECIFICATIONS>
Ratings; 100-240VAC
Acceptablerange; 90to264VAC
Ratings; 50/60Hz
Acceptablerange; 50Hz±5%,60Hz±5%
Power Consumption;
15VA
SC450G-D:
Ratings; 12-24VDC
Acceptablerange; 10.8to26.4VDC
Power Consumption;
10 W
Safety, EMC and RoHS conforming standards
Safety: EN61010-1
EN 61010-2-030
EN 61010-2-201
CAN/CSAC22.2No.61010-1
CAN/CSAC22.2No.61010-2-030
CAN/CSAIEC61010-2-201
UL61010-1
UL61010-2-030
UL61010-2-201
EMC: EN61326-1ClassA,Table2
*1
EN 61326-2-3
EN 61000-3-2 Class A
EN 61000-3-3
RCM:EN61326-1ClassA
KoreaElectromagneticConformityStandard
*1: Inuenceofimmunityenvironment(CriteriaA):Outputshiftisspeciedwithin±35%ofF.S.
Note: ThisinstrumentisaClassAproduct,anditisdesignedforuseintheindustrialenvironment.
Please use this instrument in the industrial environment only.
RoHS: EN50581
FMnonincendiveapproval(sufxcodeType:-U):
FM3611ClassI,Div.2,GroupABCD,T6forTa-20to55ºC
Installationaltitude:2000morless
CategorybasedonIEC61010:II(Note)
PollutiondegreebasedonIEC61010:2(Note)
Note: Installationcategory,calledover-voltagecategory,speciesimpulsewithstandvoltage.Category
II is for electrical equipment. Pollution degree indicates the degree of existence of solid, liquid,
gas or other inclusions which may reduce dielectric strength. Degree 2 is the normal indoor
environment.
Environment and operational conditions:
Ambient temperature;
-20 to +55ºC
Storage temperature;
-30to+70ºC
Humidity; 10to90%RH(non-condensing)
Dataprotection; EEPROMforconfigurationdataandlogbook.Lithiumcellfor
clock.
Watchdog timer; Checks microprocessor
Power down; Reset to measurement.
Automatic safeguard;
Auto return to measuring mode when touchscreen is untouched
for 10 min.