IM 12D08N05-01E
APP-4
<APPENDIXES>
n
Temperaturecompensationmatrix.
1. A minimum number of values is required to make interpolation possible. The
highlighted values marked as are mandatory to enter.
Concent Tref T1 .... Tx .... T10
Sol1 C1 S1Tr S1T1 S1T10
....
Solx SxTx
....
Sol10 C10 S10Tr S10T1 S10T10
2. Tref(referencetemperature)isdefinedintheTemperatureCompensationmenu.If
TrefisbetweenT1andT10thenthevalueofTrefneedstobeenteredasTx(T2....T9).
Concent Tref T1 .... Tx .... T10
Sol1 C1 S1Tr S1T1 S1T10
....
Solx
....
Sol10 C10 S10Tr S10T1 S10T10
3. ForeverySxTxthatisenteredthefollowingvaluesbecomemandatorytoenter:Cx,
SxTr, SxT1, SxT10 and Tx.
Concent Tref T1 .... Tx .... T10
Sol1 C1 S1Tr S1T1 S1T10
....
Solx Cx SxTr SxT1 SxTx SxT10
....
Sol10 C10 S10Tr S10T1 S10T10
The matrix can be cleared before
entering new values. Next new matrix
values can be entered as described
above. The SC450G can interpolate the
matrix. During this process it will check
ifthematrixiscompletelyascending/
descending. This is necessary as
otherwise the lookup function can give
two results for one temperature. If an
error is found, the SC450G will specify
the location of the error as shown in the
userinterfacescreen“userdened1/2”.
The backspace key should be used for
deleting an individual matrix value.
An empty value is shown as
.