IM 12D08N05-01E
APP-6
<APPENDIXES>
Sensordamage,and/orcoatingscanbedifficulttoseeandthecalibrationcheckcan
conrmtheirpresence,byadeviationfromtheknownsolutionconductivity.Theremedial
actionshouldbetocleanthesensor,andcarefullycheckforblockageordamage(not
simplytorecalibrate).
Higher conductivity solutions should be used where possible. The lower the conductivity of
the test solution, the easier it is to contaminate. Carbon dioxide from the air can be quickly
absorbed to cause an error. All containers must be suitably clean, and all materials suitably
pure.Outsideofawell-equippedlaboratorytheseconditionsarehardtomeet.
Also note that the check must be performed with due regard to the cell constant of the
sensor, as it limits the effective working range. The documentation provided with the sensor
shouldbeconsultedtodetermineitsmaximumworkingvalue.Byusingacheckingsolution
close to the upper range limit of the sensor, the detection of contamination is better. A
slightlycontaminatedsensormayreadperfectlyatlowerconductivity,butshowsignicant
errors at higher values. This is caused by the early onset of polarization, the factor that
limits the upper range of the sensor in any case. Polarization is seen in contaminated
sensors, and it is for this reason that there is a sophisticated polarization check built into the
SC450G.Detailsofthisdiagnostictoolarefoundinthetroubleshootingsection(Chapter8).
SC450G is programmed with the following table of conductivity of Potassium Chloride
(KCl)solutionsat25°C.ThisisusedintheAutomaticCellConstantsettingcalibration
feature.(Seechapter6oncalibration)ThetableisderivedfromtheStandardslaiddown
in“InternationalRecommendationNo.56oftheOrganisationInternationaledeMétrologie
Legale”.
Table 2 KCl values at 25 °C
mol/l mg KCl/kg Conductivity
0.001 74.66 0.1469mS/cm
0.002 149.32 0.2916mS/cm
0.005 373.29 0.7182mS/cm
0.01 745.263 1.4083mS/cm
0.1 7419.13 12.852mS/cm
1.0 71135.2 111.31mS/cm
Ifitismoreconvenient,theusermaymakesolutionsfromSodiumChloride(NaCl)withthe
helpofthefollowingrelationshiptable.ThistableisderivedfromtheIECnorm60746-3.