IM 12D03D02-01E
1-3
1. Outline
Battery life: Approximately 200 hours*
4
of continuous use (battery type
and operating condition dependent)
Functions: Data memory (300 points), alarm clock
EMC Compliance:
EMI (Emission): EN 61326-1 Class B
Test Item Frequency Range Basic Standard
Electromagnetic radiation disturbance 30 to 1000 MHz CISPR 16-1 and 16-2
EMS (Immunity): EN 61326-1 Table 2 (For ise in industrial locations *
5
)
No. Test Item Test Specication Basic Standard
Performance
Criteria*
1
Electrostatic
discharge
4 kV (contact)
8 kV (air)
IEC 61000-4-2 A
2
RF amplitude
modulated
electromagnetic eld
80 to 1000 MHz,
10 V/m (unmodulated)
80% AM (1 kHz)
IEC 61000-4-3
B**
1.4 to 2.0 GHz,
10 V/m (unmodulated)
80% AM (1 kHz)
A
2.0 to 4.0 GHz,
3 V/m (unmodulated)
80% AM (1 kHz)
A
* A: Normal performance within the specication limits: ±20% of the measured value.
B: Temporary degration or less of function or performance which is self-recoverable.
** Display value may be a󰀨ected by strong electromagnetic eld.
EMC Regulatory Arrangement in Australia and New Zealand (RCM)
EN 61326-1 Class B, Table2
Korea Electromagnetic Conformity Standard Class B
한국 전자파적합성 기준
B급 기기 (가정용 방송통신기자재)
 이 기기는 가정용(B급) 전자파적합기기로서
주로 가정에서 사용하는 것을 목적으로 하며,
모든 지역에서 사용할 수 있습니다.
Environmental resistance:
Compliant with RoHS*
6
, WEEE, and EU battery directive
*1: Resistivity can be measured for reference. In that case repeatability is determined by
conductivity.
*2: Display range is from-10 to 120°C. 0 to 50ºC when the sensor cable is immersed in water.
*3: When measuring range is set to full scale.
*4: When alkaline batteries are used.
*5: Display value may be a󰀨ected by strong electromagnetic eld.
*6: RoHS: EN IEC 63000 (Style 2.03 and later)