IM 12D03D02-01E
2-5
2. Preparation
2.6 Setting Temperature Compensation Coe󰀩cient
As described in Section 7.3, liquid conductivity varies with liquid temperature.
Therefore, if concentration is measured by conductivity, the conductivity must be
converted to equivalent conductivity at a certain temperature. This instrument
incorporates standard temperature conversion functions to convert liquid conductivity
measurements to equivalent conductivity at 25°C. To display equivalent liquid
conductivity at 25°C, set the temperature compensation coe󰀩cient as described in
this section.
Temperature coe󰀩cient for NaCl solutions is stored in this instrument. If any other
solution is used, set the temperature compensation coe󰀩cient manually.
Refer to Sec. 4.3 (2) Temperature compensation setting (t.Co) panel.