2. Preparation
2.6 Setting Temperature Compensation Coecient
As described in Section 7.3, liquid conductivity varies with liquid temperature.
Therefore, if concentration is measured by conductivity, the conductivity must be
converted to equivalent conductivity at a certain temperature. This instrument
incorporates standard temperature conversion functions to convert liquid conductivity
measurements to equivalent conductivity at 25°C. To display equivalent liquid
conductivity at 25°C, set the temperature compensation coecient as described in
this section.
Temperature coecient for NaCl solutions is stored in this instrument. If any other
solution is used, set the temperature compensation coecient manually.
Refer to Sec. 4.3 (2) Temperature compensation setting (t.Co) panel.