5. Maintenance
5. Maintenance
5.1 For Optimum Meter Performance
The SC72 meter appears to be very simple, but is a precision instrument. To ensure
that measurement accuracy is maintained, observe the following precautions
regarding preliminary setting, measurement, maintenance and storage.
Preliminary setting, measurement, maintenance and storage:
Measurement
Maintenance
Preliminary
setting
Storage
Table 5.1 Observe the following precautions
Preliminary
setting
Setting cell constant
Set specic electrode cell constant.
Setting temperature coe󰀩cient
Set liquid temperature coe󰀩cient when the standard temperature
conversion is required (see Section 2.5)
If standard temperature conversion is not requiresd, set 0.00.
Measurement
Select one of the three conductivity measurement ranges (0 to
200µS/cm, 0 to 200mS/cm, or 0 to 2S/cm) corresponding to the
type of electrode used.
Temperature measuring range is 0 to 80°C.
Maintenance
When measurement is completed, thoughly wash electrode to
remove the measured liquid.
If the electrode gets stained or deformed, its cell constant may
change. In this case, calibrate the meter with standard solution
(see Section 5.4).
Storage
Avoid high-temperature, high-humidity storage locations.