IM 12D03D02-01E
5-3
5. Maintenance
5.3 Cleaning and Drying Connectors
Deteriorated insulation between connector pins can cause inaccurate readings. To
remove stains and/or moisture that may cause deteriorated insulation, clean the
connector with a dry cloth or a cloth moistened with anhydrous alcohol. If necessary,
use a dryer.
Wipe off stains and/or moisture on
meter connector with a dry cloth.
Check that there are no stains on
the O-ring.
O-ring
Hot air
Use a hair drier if necessary to
remove moisture from
connector of sensor cable.
CAUTION
Use anhydrous alcohol to clean the connectors so no moisture remains. Dry
connectors completely.