IM 12D03D02-01E
6-3
6. Troubleshooting
(5) or Out of measurement range
Possible Cause:
Conductivity is over the maximum value on the range.
Corrective Action:
For manual ranging, select an appropriate range. If value exceeds maximum of
highest range (or when on autorange), use a more appropriate sensor (see Sec. 4.3
(3)).
6.3 Causes of Abnormal Measured Value
If error messages do not occur, but measured values seem incorrect, check the
following:
Are cell constants and temperature coe󰀩cients correctly set?
Is the sensor properly connected to the meter?
Are bubbles attached to the electrode portion?
Is the sensor damaged or dirty?
6.4 Other Conditions
An alarm sounds
The alarm is set to sound at the preset alarm time. Refer to Section 4.3 (9), “Alarm
time setting (ALM) panel.”
• Beeps
The beep sound to acknowledge a key press can be enabled/disabled. Refer to
Section 4.3 (11), “Set beep on/o󰀨 (bZ.o) panel.”