App-12
IM DL850E-01EN
Notes about Executing FFT Computation
Computation is normally performed on the sampled data in the acquisition memory. However, for waveforms
that have been acquired in envelope mode, computation is performed on the maximum and minimum values per
acquisition interval.
Appendix