7-34
IM WT3001E-51EN
Setting the Display Range of the Bar Graph
Set the display range of the bar graph using harmonic order.
The display range of bar graph 1 to bar graph 3 is the same.
The minimum value is 0
th
order (DC). However, if the measurement function is set
toφ,φU,orφI,thereisnovalueforthe0
th
order. Thus, 0
th
order is not displayed on
the bar graph.
• IfthemeasurementfunctionissettoφUorφI,thereisnovalueforthe1
st
order.
Thus, 1
st
order is not displayed on the bar graph.
The maximum setting is 100
th
order. However, bar graphs of orders that exceed the
upper limit of the measured order (see section 7.9) are not displayed.
Cursor Measurement
• ON/OFF
A cursor can be placed on the displayed bar graph to measure the value at that
point.
ON: Performs cursor measurement.
OFF: Does not perform cursor measurement.
Measurement Item
Y1+: Y-axis value of cursor + of bar graph 1
Y1x: Y-axis value of cursor x of bar graph 1
• ∆Y1:ThedifferencebetweentheY-axisvaluesofcursor+andcursorxofbar
graph 1
Y2+: Y-axis value of cursor + of bar graph 2
Y2x: Y-axis value of cursor x of bar graph 2
• ∆Y2:ThedifferencebetweentheY-axisvaluesofcursor+andcursorxofbar
graph 2
Y3+: Y-axis value of cursor + of bar graph 3
Y3x: Y-axis value of cursor x of bar graph 3
• ∆Y3:ThedifferencebetweentheY-axisvaluesofcursor+andcursorxof
bar graph 3
Moving the Cursor
Two cursors (+ and x) are displayed in each graph (bar graphs 1 to 3).
The cursor position is set using harmonic order.
Bar graph displays the harmonic order indicating the cursor position.
The position of cursor + is displayed as in order+: 2.
The position of cursor x is displayed as in orderx: 55.
The harmonic order indicating the positions of cursor + and x are common to bar
graph 1 to bar graph 3.
Note
If immeasurable data exists, *** is displayed in the measured value display area.
Limitations on the Bar Graph Display by Measurement Modes
Bar graph of harmonic measurement cannot be displayed in the following
measurement modes.
IEC harmonic measurement mode
Waveform computation mode
FFT mode
Voltage fluctuation and flicker measurement mode
Cycle-by-cycle measurement mode
7.9 Displaying the Bar Graph and Making Cursor Measurements