3-6
IM WT3001E-51EN
Explanation
The numeric data can be output using ± 5-V FS DC voltage from the D/A output
connector on the rear panel. Up to 20 items (channels) can be specified.
CAUTION
Do not short the D/A output terminal or apply external voltage to it. If you do, the
instrument may malfunction.
French
ATTENTION
Ne pas court-circuiter la borne de sortie D/A et ne pas y appliquer de tension
externe. Le cas échéant, un dysfonctionnement de l’instrument est possible.
Limitations on the D/A Output Function by Measurement Modes
There are limitations on the setting and execution of the D/A output in measurement
modes other than normal measurement mode.
There is a limitation on the numeric data that can be measured and delivered through
D/A output in the measurement modes listed below. For details on the numeric data
limitations, see appendix 11 in the User’s Manual IM WT3001E-01EN. This instrument
outputs 0 V where no numeric data exists.
Wide bandwidth harmonic measurement mode
Waveform computation mode
FFT mode
The D/A output function cannot be used in the following measurement modes. This
instrument outputs 0 V on all channels.
IEC harmonic measurement mode
Voltage fluctuation and flicker measurement mode
Cycle-by-cycle measurement mode
D/A Output Item
Selecting the Measurement Function
The types of measurement functions that can be selected are the items that are
indicated in “Types of Measurement Functions during Normal Measurement”
of section 2.2, “User-Defined Function” and “Corrected Power” of section 2.5,
“Measurement Functions of Integration” of section 2.6 in the User’s Manual IM
WT3001E-01EN, as well as “Motor Evaluation Function (Option)” of chapter 1, “Delta
Computation” of chapter 6, and “Harmonic Measurement in Normal Measurement
Mode (Option)” of chapter 7 in this manual.
You can also select not to output the measurement functions (None). For channels
that are set to None, 0 V is output since no corresponding data exists.
If the range mode (see section 3.3) is set to fixed, the D/A output of a channel
whose measurement function is set to Z, Rs, Xs, Rp, Xp, or F1 to F20 is fixed to 0 V.
It is output when the range mode is set to manual.
3.2 Setting the D/A Output