3.8 Touchscreen Calibration
The touchscreen is calibrated at the factory and should not require a calibration upon installation. If the
touchscreen functionality is lost, the touchscreen may require calibration.
1. Select the Screen tab.
2. Under the Screen Calibration heading, click on the Calibrate Screen button.
Figure 3-10
3. The screen will change to a solid white screen. A message will appear that states the touchscreen
calibration will begin. The user will have the option to abort the calibration, if necessary.
4. A black circle will appear on the screen. The user will be asked to touch the inside of the circle. Touch
the inside of the circle with a stylus. The circle must be touched within 30 seconds or the calibration will be
aborted.
5. The user will be guided to repeat the touch calibration in three different locations on the touchscreen (4
touches total).
6. When the calibration is complete, the screen will return to the Screen tab window.