52
7. Select Add Cycle to set the parameters for the sampling cycles. A section will be added to the existing
window which allows the user to program the first cycle of the profile.
8. Enter the Cycle Time. If only one cycle will be used for the profile, the time will comprise the duration of the
complete profile. If multiple cycles will be used for the profile, then the time entered will be the duration of that
cycle only.
9. Enter the Sampling Frequency in minutes. Ensure enough time is allocated for sampling all vessels using
the profile. Cycle times typically range from 3 to 10 minutes, depending on the sample size, pump speed and
cycles used. A sampling frequency value of 0 is used for continuous sampling.
10. Select Settings to program the cycle Sampling Volume. The Settings window will appear.
11. Enter the Sample Volume. The sample volume includes both the purge and analytical sample
volumes.
12. Select Save to set the cycle parameters.
13. If no more cycles will be used, select Save and then select Back. The Scheduler window will
reappear.
If additional cycles will be needed, proceed to Step 14.