Floor load on the supports (A)
CONTURA G2
X=700, Y=700 34.8 N/cm
2
X=700, Y=1000 43.1 N/cm
2
CONTURA G2
X=1000, Y=1200
65.4 N/cm
2
X=1000, Y=1600
81.5 N/cm
2
X=1000, Y=2100
132.1 N/cm
2
CONTURA G2 w. granite U-table (option)
X=1000, Y=1200
70.6 N/cm
2
X=1000, Y=1600
88.2 N/cm
2
X=1000, Y=2100
143.1 N/cm
2
Support (A): Cast iron plate Ø 150
(included in delivery)
4.4 Arrangement of the Steel Plates for Contura G2
The relative unevenness of the bearing surfaces must not exceed max. 25mm (1“)..
The deviation tolerance for the center-to-center plate distance is (A) ±10 mm (all dimensions in mm)
Point of support (tilt protection only)