Installation Instructions
consult a statics engineer who is familiar with your local
floor conditions. To determine the floor load exerted by
the CMM, see the pertaining information in this bro-
chure.
0,1
1
10
100
1000
0
10 20 30 40 50
X axis horizontal
Y axis horizontal
Z axis vertical
Permissible foundation acceleration for
Contura G2 w. trigger probe and standard damping
Acceleration values above the curve
require an additional foundation
Acceleration [mm/ss]
Frequency [Hz]