EasyManua.ls Logo

Zeiss CONTURA G2 - 4.2 Installation Dimensions for CONTURA G2 with opt. U-profile

Zeiss CONTURA G2
20 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page
10
Installation Instructions
a
c
850
min. 800
min. 800
203
283
1048
T- T
t
2403
b
CONTURA G2
2653
3003
c
d
e
f
g
4.2 Installation Dimensions for CONTURA G2 with opt. U-profile (all dimensions in mm, scale 1:30)
CONTURA G2 with optional U-profile
X/Y/Z
Measuring range
Net weight of
CMM
Max. permissible
workpiece weight
1000/1200/600 ca. 2583 kg 1150 kg
1000/1600/600 ca. 3163 kg 1500 kg
1000/2100/600 ca. 5755 kg 1814 kg
c
= On-site air supply connection, precleaned compressed air
supply with pressure of 6-10 bar.
d = Power supply (two outlets with max. 16 A) to be installed
by the customer.
e = Telephone connection (ISDN)
f = CMM compressed air supply connection
g =CMM control connecting cable
t = Transport height without base
CONTURA G2 with optional U-profile
X/Y/Z
Measuring range
Length
- a -
Length
- b -
Length
- c -
Length
- t-
1000/1200/600 1743 2030
1200
1806
1000/1600/600 1743 2430
1000/2100/600 1743 2945 1959
Height of
secured CMM
assembly without
pallet and
without Z tower
Min. overhead clearance: 200 mm
Controller
Length: ca. 756
Width: ca. 575
Height: ca. 700
T-T = distance betw. table
and lower edge of probe
VAST
XT
probe 919
RDS probe 868
Height of
secured
CMM assy
incl. pallet
CONTURA G2
X/Y/Z
Measuring range
Max. cable leng.
CMM control
g
1000/1200/600 ca. 6300 mm
1000/1600/600 ca. 6000 mm
1000/2100/600 ca. 5700 mm