series User Manual 2660021145640 A
Testing
5-17
3 Enter patient data. A patient name, date of birth, and Patient ID is
required for all of the following: saving to disk, STATPAC calculations,
automatic trial lens calculations, and screening tests using Age
Corrected mode.
4 Change test parameters. Select the test parameters and the
fixation monitoring system to best suit your patient’s needs.
5 Patch the non-test eye. Reduce the room illumination. Give the
patient test instructions. Adjust the table and perimeter to a
comfortable height for the patient. Make sure the patient is sitting
comfortably. Refer to “Preparing the Patient,” on page 3-24 for full
details of patient preparation. You also may r
efer to the Help Menus by
pressing the
HELP
icon (question mark icon) on the right-hand side of
the screen.
6 Press START.
7 If the Foveal Threshold parameter is turned on, it will activate now. Refer to the discussion of
“Foveal Threshold,” on page 5-7.
8 If the fixation monitoring parameter is set to “Gaze Track” or “Gaze/Blind Spot”, follow the
on-screen instructions for setup. Refer to “Gaze Tracking (Models 740i, 745i, 750i),” on
page 5-4 for details. After gaze monitor initialization, a pop-up window appears and pr
ompts
you to start the test. Remember to monitor the patien
t during the visual field test to insure
accurate results.