Humphrey Field Analyzer II-
i
series User Manual 2660021145640 A
STATPAC Analysis & Printing
7-17
the screening printout for both eyes, screening tests may be central or peripheral but not full field
patterns (see Figure 7.7). The printout of both eyes is also known as the “O. U. Printout”.
The Quantify Defects Full Field screening test is pri
nted on two pages. The first page consists of the
Full Field printout. The second page is an enhanced view of the Central 30 degrees, allowing for
easier reading of the central portion of the printout. If there are no defects quantified in the Central
30 degrees, the printout will only be one page long. The printout will be in the full field format.
Screening Test Printouts
The type of test and test parameters are printed at the top of the printout along with the patient
d
ata, test date and test time. Like threshold test printouts, screening printouts include reliability
indices to help you determine the reliability of the patient’s responses.
When a screening test uses the Threshold Related testing mode, the cen
tral (and peripheral)
reference level values are determined from patient responses and appear on the printout and test
screen. When the Age Corrected mode is used, the central (and peripheral) reference levels display
values based on the patient’s age.