File Functions
Humphrey Field Analyzer II-
i
series User Manual 2660021145640 A
10-22
How to Transfer Tests from the HFA I to the HFA II-
i
Series via Serial Cable
1.
Connect the HFA I and HFA II-
i
series perimeters with an HFA I to HFA II-
i
Serial Transfer Cable
(P/N 52416). Plug the end of the cable with nine (9) pins into the Data Transfer port on the HFA
II-
i
(refer to diagram in Chapter (1), "Introduction & Instrument Setup," for port location). Plug
the 25 pin connector into the serial port on the back of the HFA I. Refer to “Additional
Co
mponents” in Chapter 1 of this manual for a detailed illustration of the ports on the HFA
II-
i
series.
2. Turn both instruments ON.
On the HFA II-
i
Series:
1. From the System Setup screen, select COMMUNICATIONS SETUP.
2. Select RS-232 SETUP.
3. On the RS-232 Setup screen, set the “Receive Settings” for BAUD RATE to 9600, PARITY to
E
v
en, DATA BITS to 7, and STOP BITS to 1.
4. Press PROCEED. The Communications Setup screen will then appear. Press DONE.
5. Press the
FILE FUNCTION
S icon on the right of the screen.
6. Select TRANSFER TESTS.
7. Select CLASSIC SERIAL as the Source.
8. Choose HARD DRIVE or FLOPPY as the
Destination, then pr
ess PROCEED.
Note: You do not need to select anythi
ng for “Export Format.”
On the HFA I:
1. From the Main Menu, select CONFIGURA
T
ION MENU.
2. Make sure the BAUD RATE is set to 9600 and the PARITY is set to EVEN on the Configuration
Me
nu screen, then select RETURN (to the Main Menu).
3. Select DISK FUNCTIONS.
4. Select TRANSMIT FILES.
5. Select the disk drive on which the tests
you want to transfer are stored.
6. If you want to transmit a limited number of files, highlight those files an
d choose SELECTION
COMPLETE. If you want to transmit all the files on the HFA I disk, do not highlight any files.
Just choose SELECTION COMPLETE.
On the HFA II-
i
Series:
1. A pop-up window will update the user on the progr
ess and success of test transfer.
2. When the “tests transferred” count on the pop-up window has not changed for sever
al sec-
onds, test transfer is complete. Press CANCEL. If you do not pr
ess CANCEL within about 10
minutes, the HFA II-
i
will automatically proceed to the next step.
3. The transferred tests will then be moved from temporary to permanen
t storage. Press OK when
the pop-up window informs you that the transmission is complete.
How to Transfer Tests from an Older HFA II to the HFA II-
i
via Serial Cable
1. Connect the older HFA II and HFA II-
i
series unit with an HFA II to HFA II-
i
Serial Transfer Cable
(P/N 52417). Plug one end of the cable into the Data Transfer port on the HFA II-
i
series (refer
to diagram in Chapter (1), "Introduction & Instrument Setup," for the port location). Plug the
other 9 pin connector into the Serial port #1 (the first 9 pin serial
connector on the left) on the
back of the older HFA II. Refer to the
HFA II User’s Manual
for a detailed illustration of the
ports on the HFA II.
2. Turn both instruments ON and wait until you see the Main Menu screen.