Humphrey Field Analyzer II-
i
series User Manual 2660021145640 A
Kinetic Testing
13-33
2For ease of use, set the HFA II-
i
for cursor use. Press USE CURSOR if
in keypad mode. Move the cursor to the area at which you wish to
test. In this example the point to be tested is located at:
Meridian: 20
Radius: 30
Note: Ther
e is no visual cue on the screen as to the location of
the static point to be tested.
Note: The point coordinates can also be input using the
Keypad option. See previous sections for entering Meridian and
Radius values with the keypad.
3Press SELECT POINT to display the Static point stimulus to the
patient.
4If the patient sees the stimulus, a small diamond will display at that
point on the screen.
The words “Not Seen” will display briefly if the patient does not see
the stimulus.
Additional static points may be checked. Only the most recent point
seen will display.
Note: All static points tested, both “Seen” and “Not Seen,” are
recorded on the Numerical Values printout.