Zeiss Humphrey II-i series - Setting Transfer Destination Drop-Down Box to EMR;PMS Host; Setting Transfer Destination Drop-Down Box to DICOM Archive

Zeiss Humphrey II-i series
580 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Networking
Humphrey Field Analyzer II-
i
series User Manual 2660021145640 A
14-42
Figure 14.46 Setting Transfer Destination Drop-Down Box to EMR/PMS Host
Figure 14.47 Setting Transfer Destination Drop-Down Box to DICOM Archive
Note: The Data Export Host or ERM/PMS Host options in the Transfer Destination drop-down box
are only available in the listing if you have licensed the HFA-NET Pro networking software on your
HFA II-
i
. The DICOM Archive option in the Transfer Destination drop-down box is only available in
the listing if you have licensed the DICOM Gateway 2.0 software on your HFA II-
i
. Otherwise, these
options will not appear in the Transfer Destination drop-down box.
3 If you selected the EMR/PMS Host option, open the Data Format drop-down box and select
XML AND IMAGE FILES. If you selected the DICOM Archive option, REPORT will be
automatically selected. A report is an Encapsulated PDF—a DICOM formatted PDF file
transmitted via DICOM protocols. Only DICOM compatible applications can read an
Encapsulated PDF file.
4
Open the Save/Transmit Option drop-down box and select SAVE AND TRANSMIT.
5
Press PROCEED.