Scroll to the bottom of the Advanced settings section and ensure that Use simple file sharing
(Recommended) is NOT selected, as shown below (Figure G.55).
Use simple file sharing
(Recommended) is NOT
selected.
Figure G.55 Folder Options Screen – View Tab
Access Through a Router: If there is a network router between your HFA and the computer you
are attempting to connect to, you may not be able to use the RCT to configure your HFA-NET
settings.
Windows Firewall: If your file server employs Windows Fir
ewall you should verify that it accepts
access on the following NETBIOS ports: 137, 138, and 139; and that it also allows access via
port 445 for CIFS over IP. See “Firewall Software,” on page G-2 for additional information.
These exceptions are exposed via the File and
Printer Sharing Service in the Windows XP
Firewall dialog Exceptions Tab. If your instrument and server are in different subnets
(example: 172. vs 10.), then you will need to change the scope to “Any”.
Non-Windows Firewall: This version was only tested with
Windows Firewall. While the
exceptions noted above may also work with a non-Windows firewall, you should consult its
documentation for information about allowing CIFS access from an IP device.
Working with a VPN: If you are operating with a VPN on the server, you may need to disable its
Security Po
licy. This may not be necessary, so try to operate the RCT first.
HFA-NET Pro Enabled: If you have licensed HFA-NET Pro, but ar
e not currently connected to an
Electronic Medical Records (EMR) system, Test Connection will return a “failed” message
because it expects a response from your EMR software. If no other failures are present, then
HFA networking features should still work.