Humphrey Field Analyzer II-
i
series User Manual 2660021145640 A
Setting-Up Tests
3-13
8The calculated trial lens data will automatically appear on the
Patient Data 1 screen, as is shown in the illustration to the left of this
step.
For Manual Trial Lens Entry:
1. From the Patient Data 1 screen, select TRIAL LENS.
2. Choose MANUAL TRIAL LENS INPUT.
3. Repeat Steps 3-6 in the preceding process.
4. Choose ENTRY COMPLETE. The manually entered trial lens data is automatically
entered on
the Patient Data 1 screen.
Note: For guidelines on selecting the proper trial lens (for manual input), r
efer to “Using Trial
Lenses,” on page 3-20 and “Examples of Trial Lens Correction,” on page 3-21.
Clearing Patient Data
Often you will want to enter information for a new
patient on a blank Patient Data screen. To
remove all information on the Patient Data 1 and Patient Data 2 screens, use CLEAR PATIENT DATA.
1From the Patient Data 1 screen, choose CLEAR PATIENT DATA.
2Read the confirmation question and answer appropriately.
Note: Clearing Patient Data only deletes information from the
screen. It does not delete information from the database, if the patient
data was previously saved.