2. After initialization, click the Validation/Testing button. As Turn on/off sensors does not work currently, activate sensors according to the description
below.
3. After validation, click the Step motors menu, and then, the Turn on/off sensors button in the Validation/Sensors menu activates. As a result, the
sensors can be controlled manually.
4. Slide edge
4.1. After turning on the sensor, the electronics on the Arm is activated.
4.2. The operation of the sensor can be validated blocking the laser beam by hand
and moving it near. As a result, the displayed value must be changing. Tuning
of the sensor is described later.
5. Slide out
5.1. After turning on, the operation led lights up.
5.2. There are three states of the led: prism on, prism off, uncertain. Tuning of the
sensor is described later.