P1000 SERVICE SOFTWARE MANUAL
7. STEP MOTORS
20
4. On the Step motors low level setup page, the following functions can be found.
5. Example of manual movement (for troubleshooting):
5.1. Select motor to be moved (for example, Stage X).
5.2. Remember the Speed and Acceleration values if you want to modify them since there is no reset button. Set Speed to 30mm/s in case of
Aperture and ND filter as the default value is only an example.
List of motors to be tested,
exclamation mark shows untested
state
Speed setting
Motor
acceleration
Start motor with default settings
Defines the
maximum
of Motor
movement
slide
Motor
movement
(steps) slide
Motor movement (steps) in
negative and positive
direction
Set test start
position
Test movement step
count
Start
motor
test
Start
validation
test