3. The high-level setup window is the same as the aperture’s. Create three positions and select position one (filtered position).
4. Click the Slice motor position button to activate the motor control panel. Click the Home button.
5. Step the disk in positive direction until the filter and the objective behind are aligned (you can light the unit with a lamp from the direction of the
condenser). When done, click on the side panel to set this value and to the Slice motor position cell.
6. Repeat steps from 10.3 to 10.5 to find the other two filter positions. Click to set the values.
7. After defining all three values, start the validation process. Set Test count to the intended value or set 1 and validate the component.