6.6. Stage slide X
Setting this parameter, you can navigate the Arm. In Slide to stage area, select the position of the marked, double-wide slide and send it to
the stage area. The Stage must be in fully open position as defined in step 6.5. If not so, open it. The markings of the Stage and the granite
slab must be aligned with the bi-sectioning line on the slide in the Arm, this has already been set in step 3.3. If not so, set it now and click
on the side panel to set this value.
6.7. Slide start pos Z
This approximate value is calculated by the software. Navigate the Arm in Z and Y directions so that the slide fit between the shoes and the
support lath. The distance between the slide and the support lath must be ~0,2-0,3mm. You get the best visibility when the top side of the
slide and the top of the lath are aligned. When done, click on the side panel to set this value.