Caution!
The software version P1000v200, 2.0.0.102326 contains an error!
Before clicking Slide to stage area, save your work and exit the software. Find the MicroscopeConfiguration.xml file on the path below:
C:\\ProgramData\3DHISTECH\SideScanner\
Manually, insert the StageSlidePosStartZ parameter in subsection <StagePositions> in section <SlideLoadingUnit>. Add here the ZMotor
HWLimitMin value (or slighly more) in section <SlideLoader3DDriver>. If you accidentally skip this step, the Gripper will collide at the objective!
After this step, restart the service software and continue working.