P1000 SERVICE SOFTWARE MANUAL
16. OBJECTIVE CHANGER UNIT
92
16. Objective Changer Unit
In this setup window, the objective, aperture and ND filter positions can be synchronized. For proper operation, synchronize the settings of the
objectives first.
Note!
The Objective change position value is identical with the objective position already set in Objective Changer!
1. Click Objective changer option in the Sub-assembly column on the left and, on Validation page, click the Focus motor position button. Send the
actuator to this position by clicking Go to II.
2. In Objective changer unit settings, click Objective change position. Save the value.