P1000 SERVICE SOFTWARE MANUAL
16. OBJECTIVE CHANGER UNIT
SERVICE GUIDE
93
Caution!
The set value overrides the value set for Focus motor position!
3. Select the objective you want to setup under the Objective selection menu. Adjust each of the objectives based on the following table. These
settings include basic values that must be fine-tuned later.
Aperture pos.
Objective
ND filter pos
Pos 1
20x
Pos 1
Pos 2
40x
Pos 2
Pos 2
40x immersion
Pos 2
4. Go to Validation page, then validate the units.