Operation A-dec 500 Assistant’s Instrumentation
Page 4 86.0603.00
Autoclavable HVE and Saliva Ejector Asepsis
The table shows the steps for HVE and saliva ejector asepsis.
Table 1 Steps for HVE and Saliva Ejector Asepsis
How to Sterilize the HVE Tip
It is recommended that the stainless steel tips used on the HVE be heat sterilized between patients using
either sterilizers of the following methods: steam autoclave, or chemical vapor (275ºF [135ºC] maximum
temperature; four minutes at temperature). If you are using disposable HVE tips, be sure to replace
them with new tips between patients.
The table shows the steps for sterilizing the HVE tip.
Table 2 Steps for Sterilizing the HVE Tip
NOTE: A-dec HVE Cannula Opening: Standard A-dec HVE – 0.435" ± 0.006" (11.05 ± 0.15 mm). A-dec
15 mm HVE – .592" (14.8 mm). If not using A-dec HVE Tips, select a tip that is compatible with
your HVE cannula opening.
A-dec Saliva Ejector Cannula Opening: A-dec Saliva Ejector – 0.265 ± 0.006" (6.73 ± 0.15 mm)
Step Action
1
Remove the tip from the valve body.
NOTE: Turn OFF the central vacuum before disconnecting the HVE or saliva ejector.
2
To remove the HVE or saliva ejector body, separate the body from the tubing by pulling it apart
at the tailpiece.
3
Remove the control valve by pushing it out of the HVE or saliva ejector valve body.
4
Clean and rinse the valve body and control valve using a mild detergent, water, and the brushes
that have been provided. Allow the instruments to dry completely.
5
Heat sterilize the valve body and control valve for 6 minutes using either of the following
methods: steam autoclave, or chemical vapor (275ºF [135ºC] maximum temperature).
6
Apply a light coat of A-dec silicone lubricant on the O-ring seals of the control valve.
CAUTION: Use only silicone lubricant when lubricating instrumentation O-rings.
Petroleum products will cause permanent damage to the O-rings.
7
Reinstall the control valve in the valve body, and the body on the tubing tailpiece.
8
Operate the HVE and the saliva ejector valves several times to verify that they rotate smoothly.
Step Action
1
Remove the HVE tip from the HVE valve body.
2
Clean and rinse the HVE tip using a mild detergent and water, then allow the tip to
completely dry.
3
Sterilize the HVE tip using one of the recommended methods.

Other manuals for A-Dec 500