Viscosity Measurement Range | 0.3 to 10, 000 mPa·s |
---|---|
Viscosity Accuracy | ±1% of reading |
Viscosity Repeatability | ±0.5% of reading |
Temperature Accuracy | ±0.1°C |
Temperature Resolution | 0.01°C |
Interface | RS-232C |
Power Supply | 100 to 240 VAC (50/60 Hz) |
Weight | 7 kg |
Step-by-step instructions for connecting the display unit to the main unit and powering the device using the AC adapter.
Guidelines for installing the viscometer in an environment free from excessive temperature, humidity, dust, and magnetic fields.
Step-by-step guide to power on, start measurement, freeze data, and stop the measurement process.
Important considerations for calibration, including temperature, sample surface adjustment, and unit settings.
Details on calculating correction values using kinetic viscosity and viscosity, and examples for one-point and two-point calibration.
Step-by-step instructions for performing a one-point viscosity calibration using a standard fluid.
Detailed procedure for performing a two-point viscosity calibration using two standard fluids.
Instructions for simplified calibration using purified water, specific to the SV-10 model.
Troubleshooting steps for unstable readings, including environment checks, vibration, and condition settings.
Checks and solutions for incorrect measurement values, such as sample surface adjustment, plate cleanliness, and bending.