B
bethany59Aug 10, 2025
Why are the measurement values not correct on my A&D SV-1A?
- MMaurice VegaAug 10, 2025
Several factors can influence the accuracy of measurement values. Here are a few things to check: * Ensure the data hold mode is off by pressing the HOLD key. * Adjust the table height to align the sensor plates with the sample surface. * Level the viscometer to ensure a leveled liquid surface. * Clean any residual sample material from the sensor plates with alcohol. * Maintain the sample surface level during measurement. * Confirm that the serial numbers of the main and display units match. * Consider periodic calibration with a standard viscosity fluid. * If using a cup, calibrate with it before measuring viscosity. * Remember that sample viscosity is temperature-dependent.