EasyManua.ls Logo

A&D SV-1A - 6 Measurement; Preparing the Sample (for the SV-A Series)

A&D SV-1A
102 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
29
6. MEASUREMENT
6-1 Preparing the Sample (For the SV-A series)
* Vibration or shock during transportation may cause a display value error. Before measurement,
calibrate the viscometer using the sample cup that will be used for measurement. For calibration
with purified water, refer to “8-2-3 Simplified Calibration Using Purified Water”. (Only for SV-1A/10A)
Model: SV-1A
1 Pour the sample into the cup until its surface is between the level
gauges.
- With the 2 mL sample cup, the lower level gauge indicates 1.8
mL, the upper level gauge indicates 2 mL.
- With the 45 mL sample cup, the lower level gauge indicates 35
mL, the upper level gauge indicates 45 mL.
*The glass sample cup does not have level gauges.
2 Attach the cup on the table along the guides.
- With the 2 mL sample cup, place the sample cup in the sample
cup holder (For 2 mL), and set it into the 45 mL sample cup.
* When using the sample cup holder,
select either the transparent
holder or the black holder, depending on the sample color.
3 Confirm that the protector is in the position as shown in the
figure.
Raise the lever to release the sensor unit.
4 Pinch the grips, support the front side of the sensor unit and
gently lower the sensor plates above the sample surface.
Adjust the X-Y-Z stage if necessary.
5 Lower the lever to secure the sensor unit.
6 Confirm that the positioning of the sensor plates is in center of
the sample cup by adjusting the X-Y-Z stage.
- When using the 2 mL sample cup, raise the protector.
Sample cup
(2 mL)
Sensor
plate
Protector
Level
gauges
Lever
Grips
Sensor unit
Guides
Sample cup
<2 mL>
Sample cup
holder
<For 2 mL>
Sample cup
<45 mL>

Table of Contents

Related product manuals