ABB Power Capacitors | Portable Capacitance Meter CB-2000 — User´s Guide 47
CB-2000
Problem Cause Remedy Section
No backlight.
The backlight automatically
turns off after 5 minutes.
Press any button to activate
the backlight.
4.1
The backlight is defective. Send the CB-2000 in for
service.
The internal clock shows wrong date
and time.
The CB-2000 has been
without power supply for
more than 10 minutes.
Connect a power supply and
set the date and time.
6.3
The CB-2000 hangs.
System error. Disconnect and connect
battery pack.
n/a
Battery status seems to show wrong
value.
The battery status is shown
for the rechargeable battery
pack. If batteries in the
battery holder are used, the
algorithm might not be able
to accurately calculate the
remaining battery power.
6.3
9. Trouble shooting