23
EN
Recommended separation distances batwasn portable and
mobile RF communications equipment and blood pressure
monitor
The "blood pressure monitor"is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The customer
or the user of the blood pressure monitor can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the "blood pressure
monitor" as recommended below, according to the maximum output power of
the communications equipment.
EMC Declaration
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters(m) can be estimated using
the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts(W)according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800M Hz,the separation distance for the higher
frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.