47
IS THERE A WAY TO MINIMIZE THE NUMBER OF
SENSOR CONDITIONING TESTS?
Sensor Conditioning Tests are an important process
which protects the sensor from environmental materials
& gases. They ensure test accuracy and the long-term
stability of your ACE KETOSCAN mini sensor. In fact,
this function is usually installed in breathalyzers used
by law enforcement to avoid incorrect readings. To mi-
nimize the number of Sensor Conditioning Tests, leave
your unit in an area of your home not exposed to any
smells or gas. Also, leave the top cover off of the unit
after testing.
DO THE SENSOR CONDITIONING TESTS COUNT
AMONG THE 300 TESTS THAT CAN BE PERFORMED
WITH THE ACE KETOSCAN MINI?
As mentioned in the user manual, with the ACE ACE
KETOSCAN mini you can cinduct 300 tests, then the
device must be sent to a service partner such as ACE
Instruments for a calibration or a sensor exchange. After
300 tests, a maximum of 20 more can be performed,
then no more tests are possible due to the possibility
of too high inaccuracy. The Sensor Conditioning Tests
are not added to the 300 tests possible with the inst-
rument! This means that 300 tests can be performed
in any case.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACE INSTRUMENTS ACE KETOSCAN mini and is the answer not in the manual?

ACE INSTRUMENTS ACE KETOSCAN mini Specifications

General IconGeneral
BrandACE INSTRUMENTS
ModelACE KETOSCAN mini
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals