24
The signs below the garbage bins are acronyms for the
chemical substances inside the battery:
Hg: Battery contains Mercury
Cd: Battery contains Cadmium
Pb: Battery contains Lead
Technical Data*
Sensor Fuel Cell Sensor
Accuracy +/- 0.075‰ (per mill) with reference to
1.00‰ at 25°C (per mill = default setting)
OR +/- 0.0075% BAC
(Blood Alcohol Concentration)
with reference to 0.100 % BAC at 25°C
OR +/- 0.0375mg/l with reference to
0.50mg/l at 25°C
Detection
Range
0.00 up to 4.00‰ (‰ = default setting)
OR 0.000 up to 0.400% BAC (USA)
OR 0.00 up to 2.00mg/L (UK)
Warm-up
Time
Within 15 – 20 seconds at 0.5‰.
This may vary depending on the value
of per mill / BAC which was measured.
Dimension 109 x 51 x 18mm (L x W x H)
Weight 84g (battery included)
Power
Supply
2 pieces of 1.5V AAA Alkaline batteries
(= power for up to 1000 tests)
Working
Temperature
5°C to 40°C
Storage
Temperature
0°C to 40°C
Calibration Every 12 months or after 500 tests. For a
calibration service you can send the device
to ACE Instruments (see www.ace-technik.
com).
*The technical data are subject to change without prior notice