Device Type | Negative Pressure Wound Therapy (NPWT) System |
---|---|
Operating Modes | Continuous and Intermittent |
Therapy Mode | Negative Pressure Wound Therapy |
Therapy Pressure Range | -50 mmHg to -200 mmHg |
Maximum Negative Pressure | -200 mmHg |
Pressure Settings | -75mmHg, -125mmHg, -175mmHg |
Indications | Chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), flaps and grafts |
Canister Volume | 300 ml |
Alarms | Blockage, Low Battery, Canister Full |
Noise Level | <45 dB |
Operating Temperature | 10°C to 40°C (50°F to 104°F) |
Storage Temperature | -20°C to 60°C |
Altitude Range | Up to 3, 000 meters |
Design | Portable |
Power Source | AC Power Adapter (100-240V, 50/60Hz) and Rechargeable Battery |
Keep the unit away from children, pets, pests, dust, lint, and heated surfaces. Do not operate if damaged.
Diagrams and explanations of the therapy unit, control panel, buttons, and LEDs.
Explanation of alert priorities (Low/Medium) and their indicators (tone, LED).
How to use Audio Pause and details on the Battery Low Alert and its resolution.
Details of the Battery Critical Alarm, its indicators, and resolution steps involving recharging and ensuring therapy is on.
Conditions triggering the Blockage Alert, its indicators, and steps to resolve, including checking clamps and tubing.
Conditions triggering the Blockage Alarm, its indicators, and resolution steps like checking clamps, tubing, and canister level.
Further steps for blockage alarm, including unit position, restarting, and changing canister if needed.
Conditions triggering the Leak Alarm, its indicators, and resolution steps like reinstalling the canister or finding leaks.
Steps to identify and fix leaks, including checking connectors, drape, and pad. Also, when to contact KCI or doctor.
Description of the System Alarm for internal errors, its indicators, and resolution by restarting the unit.
Emphasizes following product instructions, seeking medical advice for questions, and emergency contact procedures.
Warnings about product use, assembly, maintenance by qualified personnel, and electrical standards compliance.
Warnings on bathing, submersion, cell phones, small parts, and allergic reactions.