Figure9
Figure10
Means have been calibrated
Means calibration take effect
Means calibration has been canceled
7.4 Cancel Calibration
,9.
, 10,
.
Access the second function of the calibration menu as shown in Figure Then push the
button the screen should be shown as Figure which means the calibration has been
canceled
zero
8
Troubleshooting
ERR1 displayed: Beyond zeroing range of 2%FS
ERR2 displayed: Sensor data out of limit contact Additel Corporation
,
ERR3 displayed: Sensor damaged contact Additel Corporation,
Digital Pressure Gauge
12
Means have been calibrated