Affymetrix GeneChip Fluidics Station 450 - User Manual

Affymetrix GeneChip Fluidics Station 450
98 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
P/N 08-0092 Rev. G
April, 2006
Affymetrix
®
GeneChip
®
Fluidics Station
450/250
User’s Guide

Questions and Answers

  • M
    Michael WigginsAug 18, 2025
    What to do if Fluidics Station X does not respond on Affymetrix Laboratory Equipment?
    • E
      Eileen WilliamsAug 18, 2025
      If the Fluidics Station X does not respond, first ensure that the power is switched on. Then, verify that the correct fluidics station is designated for communication. Finally, check that all cables are firmly connected to the fluidics station.
  • M
    Matthew GibbsAug 21, 2025
    How to fix a Missing Fluid Error on Affymetrix GeneChip Fluidics Station 450?
    • J
      John PerezAug 21, 2025
      If the Affymetrix Laboratory Equipment shows a Missing Fluid Error, several issues could be the cause. Check for holes in the cartridge septa, ensure the sample or staining solution is correctly placed and contains sufficient volume (500 μL). Also, inspect for blockages in the sampling tube or fluidics station line, or a failure of the fluidics sensors. Lastly, see if the pump tubing is stretched too tightly. To resolve this, run the Recover script and use another cartridge if necessary. Ensure the sample or stain vial is properly placed. Add more sample solution if needed. Run the Clean or Prime script with fresh deionized (DI) water to clear any salt blockage. If the tubing is too tight, loosen the clamps to allow it to relax, then close them.
  • J
    Jason OwensAug 23, 2025
    What to do if I see Error While Draining or Error While Filling on Affymetrix GeneChip Fluidics Station 450 Laboratory Equipment?
    • A
      Alejandra BestAug 23, 2025
      If you encounter an Error While Draining or Filling on your Affymetrix Laboratory Equipment, it could be due to several reasons. These include defective septa in the cartridge, insufficient sample or stain volume, excessive bubbling in the cartridge, buffer conductivity being too low, or failure of one of the fluid sensors. As a solution, try using a new cartridge, adding more sample solution to the sample vial, changing the buffer to reduce detergent or increase salt.
  • N
    Nathan LewisAug 26, 2025
    What to do if I see Sensor Timeout on Affymetrix Laboratory Equipment?
    • A
      Angela MartinezAug 26, 2025
      If the Affymetrix Laboratory Equipment displays a Sensor Timeout, it means there was no user response to a prompt, such as "Remove Vial". To resolve this, simply start the selected script again.
  • A
    Autumn MonroeAug 27, 2025
    Why is buffer leaking inside the Affymetrix Laboratory Equipment fluidics station?
    • M
      Monique Neal DVMAug 28, 2025
      If buffer is leaking inside the fluidics station, this could be due to loose tubing attachments inside the fluidics station, the washblock requiring replacement or salt buildup in the lines of the fluidics station. Run the Clean or Prime script with fresh DI water to flush out salt blockage.
  • Z
    Zachary SheltonAug 30, 2025
    What causes cartridge needles of the fluidics station not engaging with the cartridge on Affymetrix Laboratory Equipment?
    • L
      Lindsay SimonAug 30, 2025
      If the cartridge needles of the fluidics station are not engaging with the cartridge, it could be due to defective septa or extra flashing on the cartridge, salt buildup on the needles, or improper alignment or engagement of the cartridge holder. Try using another cartridge. Run the Clean script with fresh DI water to flush out salt blockage. Clean cartridge needles with a wet cotton swab. Place the cartridge into the cartridge holder and firmly lift the lever to engage the cartridge needles.
  • D
    Daisy ConnerSep 1, 2025
    Why sample needles do not properly enter vial on Affymetrix Laboratory Equipment?
    • D
      dmcfarlandSep 1, 2025
      If the sample needles do not properly enter vial on Affymetrix Laboratory Equipment, it may be due to a bent sample needle or the user pressing the needle lever down too quickly or with too much force. Replace sample needle. Engage sample needle lever more slowly and/or with less force.
  • K
    kprestonSep 3, 2025
    Why I recovered less sample than initial input during Recover script on Affymetrix Laboratory Equipment?
    • J
      Jessica AlexanderSep 3, 2025
      If you recovered less sample than initial input during Recover script on Affymetrix Laboratory Equipment, it is caused by loose tubing attachments inside the fluidics station.
  • L
    Lisa PalmerSep 5, 2025
    What does Temperature Timeout mean on Affymetrix GeneChip Fluidics Station 450 Laboratory Equipment?
    • A
      Amy WhitneySep 6, 2025
      If the Affymetrix Laboratory Equipment displays a Temperature Timeout, it indicates that the temperature has not reached the required level within the expected time, assuming the ambient temperature is within the operating specifications (15 – 30 degrees C).
  • P
    phyllisnorrisSep 8, 2025
    What to do if the Affymetrix GeneChip Fluidics Station 450 displays an Improper Script error?
    • B
      Brenda BraySep 8, 2025
      If the Affymetrix Laboratory Equipment displays an Improper Script error, it means that the user is attempting to run a FS400 script on FS450Dx. To solve this, download the correct FS450Dx script and continue.