Maintenance Manual -
Revision 1.4 -
English version

Questions and Answers

  • S
    Shelley LeeAug 15, 2025
    What to do if AFi Lisa AFI-C200 Laboratory Equipment shows 'IMBALANCE'?
    • K
      kingcodyAug 15, 2025
      If your AFi Laboratory Equipment displays the 'IMBALANCE' error, it means the rotor is not loaded symmetrically or the centrifuge is installed on an unsuitable surface, or the detection threshold is set incorrectly. Here's what you should do: 1. Wait for the automatic shutdown to complete. 2. Open the lid using the button. 3. Press Start/Stop to clear the message. 4. Balance the rotor loads. 5. Place the centrifuge on a compliant surface. 6. Restart the spin cycle or calibrate the imbalance sensitivity threshold.
  • Y
    yolandaroseAug 18, 2025
    How to fix no display on AFi Lisa AFI-C200?
    • J
      Jesse JacksonAug 18, 2025
      If your AFi Laboratory Equipment screen remains black, it could be due to no voltage. Try the following: * Operate the switch under the Control Center. * Connect the mains power cable. * Check the electrical supply of the laboratory and restore it.
  • K
    krista67Aug 21, 2025
    What to do if AFi Lisa AFI-C200 displays 'MOTOR OVERTEMPER ATURE'?
    • J
      Justin WallsAug 21, 2025
      If your AFi Laboratory Equipment shows a 'MOTOR OVERTEMPER ATURE' message, it means the motor temperature is too high. A stop in “freewheel” mode occurs. Wait 30 minutes (software reset) before opening the lid. You may also need to activate the air conditioning system to lower the ambient temperature or space out the periods of use. If the motor is damaged or there's a motor temperature sensor connection malfunction, the motor should be changed.
  • S
    Samantha ValenzuelaAug 25, 2025
    What to do if AFi Lisa AFI-C200 displays 'BOWL OVERTEMPER ATURE'?
    • L
      Logan KennedyAug 25, 2025
      If your AFi Laboratory Equipment displays 'BOWL OVERTEMPER ATURE', it indicates that the temperature in the bowl has exceeded 43°C, potentially due to a high room temperature, an inappropriate temperature setpoint, or a high initial loading temperature. You can: * Wait for the automatic shutdown of the rotor to complete. * Activate the air conditioning system to ensure an ambient temperature as described in the user’s manual. * Adjust the speed/temperature settings of the program to ensure they are compatible with normal use. * Place the samples in an environment between 4 and 37°C for 1 hour before centrifugation. * Check the operation of the chiller with the menu “Auto-Diagnostic “ (Startup and performance).
  • H
    hlevyAug 28, 2025
    What to do if AFi Lisa AFI-C200 Laboratory Equipment displays 'LOCK FAIL'?
    • T
      Tina FreemanAug 28, 2025
      If your AFi Laboratory Equipment displays 'LOCK FAIL', it could be because the lid is open at startup or the locked position detector is faulty. To resolve this: * Press Start / Stop to clear the message, then close the lid and wait for it to fully close before starting a cycle. * Check that the micro contact 2 blade is present and not folded, ensure it is wired correctly, and replace it if necessary.
  • M
    Michael FordAug 31, 2025
    What to do if AFi Lisa AFI-C200 shows 'Missing rotor'?
    • Z
      Zachary ShepherdAug 31, 2025
      If your AFi Laboratory Equipment indicates a missing rotor, it means a cycle was started without a rotor inside the bowl. Switch the power off and then on again. Then, assemble a rotor.
  • A
    Andrew KingSep 3, 2025
    What to do if the lid of AFi Lisa AFI-C200 Laboratory Equipment does not close?
    • I
      izamoraSep 3, 2025
      If the lid of your AFi Laboratory Equipment does not close and the mechanism doesn't start, it's possible the lid was opened with the manual opening key. Press the control button to open the lid; this should start the mechanism, allowing you to close the lid.
  • E
    Eric KingSep 7, 2025
    What to do if AFi Laboratory Equipment speed goes down and the value is blinking?
    • N
      narnoldSep 7, 2025
      If the speed of your AFi Laboratory Equipment decreases and the value is blinking, it indicates that the cycle was interrupted after a short power outage during centrifugation, causing the rotor to slow down freely. Wait for the rotor to stop completely, then switch the power off and on.
  • S
    Sheila WhiteSep 12, 2025
    What to do if there is 'TEMPERATUR E CONTROL ISSUE' with AFi Lisa AFI-C200?
    • T
      taylortonySep 12, 2025
      If you are experiencing a 'TEMPERATUR E CONTROL ISSUE' with your AFi Laboratory Equipment, it suggests that the temperature probe may be out of service. Check that the temperature probe is correctly installed under the lid, and replace it if needed. Remove the probe and measure the resistance; the value should be 500 + or - 100 Ohm (Probe PT500).