EasyManua.ls Logo

Agilent Technologies 53131A - Page 107

Agilent Technologies 53131A
252 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 Service
About the Agilent 53131A/132A Calibration Menu
Assembly-Level Service Guide 2-9
2
To View the Calibration Menu and Security Status
The Calibration menu is accessed by holding the Scale & Offset key and cycling the
POWER key. The first item in the Calibration menu is a message indicating the
calibration security status: CAL SECURE or CAL UNSECURE.
To Unsecure for Calibration
1 Press and hold Scale & Offset key, then cycle POWER key.
CAL SECURE is displayed.
If CAL UNSECURE is displayed, then the Counter is already unsecured for
calibration, and the remaining steps are not applicable.
2 Press Scale & Offset key until CODE: is displayed.
3 Press the appropriate arrow keys to enter the security code “53131”, for example,
if 53131 was the security code set at the factory.
4 Press Enter key.
CAL UNSECURE is displayed.
NOTE If “53131” was NOT the correct security code, the Counter would have displayed
BAD CODE.
To Initiate the Calibration Routines
1 Press and hold Scale & Offset key, then cycle POWER key.
2 Unsecure for calibration by performing the preceding procedure.
3 Press Scale & Offset key until CAL: OFFS1? is displayed.
深圳德标仪器 135-1095-0799

Related product manuals