Chapter 2 Service
The Agilent 53131A/132A Calibration Procedures
Assembly-Level Service Guide 2-21
2
5 Press Enter key.
The Counter displays CALIBRATING, and then it should display TI CAL PASS.
If the fail message is displayed, refer to the troubleshooting section in this chapter.
6 Disconnect the calibration setup.
This completes the TI QUIK? calibration procedures.
Depending on which timebase the Counter contains, proceed to the procedure “To
Calibrate the Standard Timebase” (page 2-23) or the “To Calibrate the High Stability
Timebase Option (Medium, High, or Ultra-High)” procedure (page 2-24) to calibrate
the timebase.
CAL: TI FINE? Calibration
1 Connect the Agilent 8130A Pulse Generator, Agilent 59992A J06 Time Interval
Calibrator, and Agilent 53131A/123A Counter as shown in Figure 2-4.
Equipment
Agilent 8130A Pulse Generator (or equivalent)
Agilent 59992A J06 Time Interval Calibrator (or equivalent)
Agilent 10503A BNC Cable (3)
Figure 2-4. Calibration Setup for CAL: TI FINE?
HP 53131A/132A
Counter
1234
In
AB
HP 59992A J06
Time Interval Calibrator
Start
Sto
p
H
P 8130A
P
ulse Generator
Output